Moku:Pro and Moku:Lab are able to switch their input impedance between 50 Ω and 1 MΩ. The input voltage range for the input impedance of each device is shown below:
50 Ω |
1 MΩ |
|
Moku:Pro | 4 Vpp |
40 Vpp |
Moku:Lab | 10 Vpp |
10 Vpp |
In the case that the 1 MΩ input impedance is enabled and the input voltage is above the 50 Ω input voltage range, then attempting to switch to the 50 Ω input impedance will be rejected by the software.
Ex: Moku:Pro has 1 MΩ enabled and a 10 Vpp signal is applied, switching to 50 Ω will not be allowed.