How to send a Laser Lock Box error signal to an output

Written by Laura Becerra

Updated at December 12th, 2025

Contact Us

If you still have questions or prefer to get help directly from an agent, please submit a request.
We’ll get back to you as soon as possible.

Please fill out the contact form below and we will reply as soon as possible.

  • Moku:Go
    Moku:Go General Moku:Go Arbitrary Waveform Generator Moku:Go Data Logger Moku:Go Digital Filter Box Moku:Go FIR Filter Builder Moku:Go Frequency Response Analyzer Moku:Go Logic Analyzer & Pattern Generator Moku:Go Oscilloscope & Voltmeter Moku:Go PID Controller Moku:Go Spectrum Analyzer Moku:Go Waveform Generator Moku:Go Power Supplies Moku:Go Lock-in Amplifier Moku:Go Time & Frequency Analyzer Moku:Go Laser Lock Box Moku:Go Phasemeter
  • Moku:Lab
    Moku:Lab General Moku:Lab Arbitrary Waveform Generator Moku:Lab Data Logger Moku:Lab Digital Filter Box Moku:Lab FIR Filter Builder Moku:Lab Frequency Response Analyzer Moku:Lab Laser Lock Box Moku:Lab Lock-in Amplifier Moku:Lab Oscilloscope Moku:Lab Phasemeter Moku:Lab PID Controller Moku:Lab Spectrum Analyzer Moku:Lab Time & Frequency Analyzer Moku:Lab Waveform Generator Moku:Lab Logic Analyzer/Pattern Generator
  • Moku:Pro
    Moku:Pro General Moku:Pro Arbitrary Waveform Generator Moku:Pro Data Logger Moku:Pro Frequency Response Analyzer Moku:Pro Oscilloscope Moku:Pro PID Controller Moku:Pro Spectrum Analyzer Moku:Pro Waveform Generator Moku:Pro Lock-in Amplifier Moku:Pro Laser Lock Box Moku:Pro Digital Filter Box Moku:Pro FIR Filter Builder Moku:Pro Phasemeter Moku:Pro Multi-instrument Mode Moku:Pro Logic Analyzer/Pattern Generator Moku:Pro Time & Frequency Analyzer
  • Python API
  • MATLAB API
  • Arbitrary Waveform Generator
  • Data Logger
  • Digital Filter Box
  • FIR Filter Builder
  • Frequency Response Analyzer
  • Laser Lock Box
  • Lock-in Amplifier
  • Oscilloscope
  • Phasemeter
  • PID Controller
  • Spectrum Analyzer
  • Time & Frequency Analyzer
  • Waveform Generator
  • Logic Analyzer & Pattern Generator
  • Multi Instrument Mode
  • Moku Cloud Compile
  • Moku general
  • LabVIEW
  • mokucli
+ More

Moku offers a Laser Lock Box (LLB) to lock a laser onto a cavity using a variety of locking methods. The output signal of the LLB is typically a control signal output from the fast and/or slow PID controllers. The error signal used for the locking process is what is passed into the PID controllers. However, there may be applications that require the output of the error signal itself.

To output the error signal, a second LLB is required to replicate the settings of the primary LLB to generate the same error signal. However, the PID controllers will be configured so that the error signal passes through to the output. Multi-Instrument mode is therefore required to deploy 2 LLBs side-by-side:

The LLB in slot 1 is the primary LLB used to lock the laser, while the LLB in slot 2 is used specifically for outputting the error signal. The input signal (Input 1) is routed to each of the LLB's main input. Output 1 is the output of the fast PID controller of the primary LLB, while Output 2 is the error signal coming from the second LLB. 

To make sure the same error signal is generated in each LLB, the demodulation sources will need to be identical as well. To do this, ensure both LLBs are setup as shown in the figure below:

 

The only difference between the two LLBs is the PID controller settings. Set the primary LLB PID controller as it would normally be configured to achieve the desired lock. In the second LLB, the fast PID controller should be set so that only the Proportional gain is enabled and set with a gain of 0 dB:

With this setup, Output 2 of the device will now output the LLB error signal.

alert output transmit laser

Was this article helpful?

Yes
No
Give feedback about this article

Related Articles

  • Can I adjust the corner frequency of the low-pass filter in the Laser Lock Box?
  • How does Moku's Laser Lock Box apply gain to the input?
  • If Moku only has two output channels, how is it possible to generate feedback control signals as well as the modulation tone and scanning waveform in Laser Lock Box?

Sitemap

  • Moku:Lab
  • Instruments
  • Software
  • Company
  • Support
  • Store
  • Terms & Conditions
  • Privacy Policy

Offices

United States
+1 (619) 332-6230
12526 High Bluff Dr
Suite 150
San Diego, CA 92130

Australia
+61 2 6171 9730
243 Northbourne Avenue
Suite 2
Lyneham, ACT 2602

Australia
+61 03 7073 3594
700 Swanston Street
Suite 5E, Level 5
Carlton, VIC 3053

Follow us

Youtube LinkedIn

官方微信

Contact us
© 2025 Liquid Instruments. All rights reserved.

Knowledge Base Software powered by Helpjuice

Definition by Author

0
0
Expand