Manually triggering waveforms simultaneously

Written by Dylan Perez

Updated at March 27th, 2026

Contact Us

If you still have questions or prefer to get help directly from an agent, please submit a request.
We’ll get back to you as soon as possible.

Please fill out the contact form below and we will reply as soon as possible.

  • Moku:Go
    Moku:Go General Moku:Go Arbitrary Waveform Generator Moku:Go Data Logger Moku:Go Digital Filter Box Moku:Go FIR Filter Builder Moku:Go Frequency Response Analyzer Moku:Go Logic Analyzer & Pattern Generator Moku:Go Oscilloscope & Voltmeter Moku:Go PID Controller Moku:Go Spectrum Analyzer Moku:Go Waveform Generator Moku:Go Power Supplies Moku:Go Lock-in Amplifier Moku:Go Time & Frequency Analyzer Moku:Go Laser Lock Box Moku:Go Phasemeter
  • Moku:Lab
    Moku:Lab General Moku:Lab Arbitrary Waveform Generator Moku:Lab Data Logger Moku:Lab Digital Filter Box Moku:Lab FIR Filter Builder Moku:Lab Frequency Response Analyzer Moku:Lab Laser Lock Box Moku:Lab Lock-in Amplifier Moku:Lab Oscilloscope Moku:Lab Phasemeter Moku:Lab PID Controller Moku:Lab Spectrum Analyzer Moku:Lab Time & Frequency Analyzer Moku:Lab Waveform Generator Moku:Lab Logic Analyzer/Pattern Generator
  • Moku:Pro
    Moku:Pro General Moku:Pro Arbitrary Waveform Generator Moku:Pro Data Logger Moku:Pro Frequency Response Analyzer Moku:Pro Oscilloscope Moku:Pro PID Controller Moku:Pro Spectrum Analyzer Moku:Pro Waveform Generator Moku:Pro Lock-in Amplifier Moku:Pro Laser Lock Box Moku:Pro Digital Filter Box Moku:Pro FIR Filter Builder Moku:Pro Phasemeter Moku:Pro Multi-instrument Mode Moku:Pro Logic Analyzer/Pattern Generator Moku:Pro Time & Frequency Analyzer
  • Python API
  • MATLAB API
  • Arbitrary Waveform Generator
  • Data Logger
  • Digital Filter Box
  • FIR Filter Builder
  • Frequency Response Analyzer
  • Laser Lock Box
  • Lock-in Amplifier
  • Oscilloscope
  • Phasemeter
  • PID Controller
  • Spectrum Analyzer
  • Time & Frequency Analyzer
  • Waveform Generator
  • Logic Analyzer & Pattern Generator
  • Multi Instrument Mode
  • Moku Cloud Compile
  • Moku general
  • LabVIEW
  • mokucli
+ More

The ability to manually trigger waveforms is available in the Moku app and API libraries. This feature allows multiple phase synchronized waveforms to be triggered simultaneously. This was previously only available with an external trigger signal. 

Moku App

Configure the waveform generator to have ‘Burst’ modulation and configure the trigger to ‘Manual’. The ‘Burst mode’ can either be set to ‘NCycle’ or ‘Start’.

Once all the necessary generators are configured, click on the settings tab and select ‘Manually trigger all channels’. This will apply the trigger signal to all waveforms configured to have a manual trigger.

Moku API (Python)

from moku.instruments import WaveformGenerator

wg=WaveformGenerator('192.168.2.200', force_connect=True)

# Configure the waveforms of each channel 
wg.generate_waveform(channel=1, type='Sine', amplitude=1, frequency=1e6)
wg.generate_waveform(channel=2, type='Square', amplitude=1, frequency=1e6, duty=50)
wg.generate_waveform(channel=3, type='Ramp', amplitude=1, frequency=1e6, symmetry=50)
wg.generate_waveform(channel=4, type='Sine', amplitude=0.5, frequency=1e6)

# Configure burst mode on each channel 
wg.set_burst_mode(channel=1, source='Manual', mode='NCycle', burst_cycles=3)
wg.set_burst_mode(channel=2, source='Manual', mode='Start')
wg.set_burst_mode(channel=3, source='Manual', mode='NCycle', burst_cycles=1)
wg.set_burst_mode(channel=4, source='Manual', mode='Start')

# Call manual_trigger() to send a trigger signal to each channel
wg.manual_trigger()


wg.relinquish_ownership()
simultaneous activation manual waveforms

Was this article helpful?

Yes
No
Give feedback about this article

Related Articles

  • How do I generate waveforms on the Moku analog outputs while running the Spectrum Analyzer?
  • Python Spectrum Analyzer 'Max Hold' function
  • How do I generate a custom filter for Moku Laser Lock Box with SciPy?

Sitemap

  • Moku:Lab
  • Instruments
  • Software
  • Company
  • Support
  • Store
  • Terms & Conditions
  • Privacy Policy

Offices

United States
+1 (619) 332-6230
12526 High Bluff Dr
Suite 150
San Diego, CA 92130

Australia
+61 2 6171 9730
243 Northbourne Avenue
Suite 2
Lyneham, ACT 2602

Australia
+61 03 7073 3594
700 Swanston Street
Suite 5E, Level 5
Carlton, VIC 3053

Follow us

Youtube LinkedIn

官方微信

Contact us
© 2025 Liquid Instruments. All rights reserved.

Knowledge Base Software powered by Helpjuice

Definition by Author

0
0
Expand