US English (US)
JP Japanese
CN Chinese
KR Korean

Contact Us

If you still have questions or prefer to get help directly from an agent, please submit a request.
We’ll get back to you as soon as possible.

Please fill out the contact form below and we will reply as soon as possible.

  • Knowledge Base Home
  • Contact Us
English (US)
US English (US)
JP Japanese
CN Chinese
KR Korean
  • Home

Why does the message, "Instrument frame rates may be reduced in this configuration" appear in Multi-Instrument Mode?

Written by Laura Becerra

Updated at April 14th, 2025

Contact Us

If you still have questions or prefer to get help directly from an agent, please submit a request.
We’ll get back to you as soon as possible.

Please fill out the contact form below and we will reply as soon as possible.

  • Moku:Go
    Moku:Go General Moku:Go Arbitrary Waveform Generator Moku:Go Data Logger Moku:Go Digital Filter Box Moku:Go FIR Filter Builder Moku:Go Frequency Response Analyzer Moku:Go Logic Analyzer & Pattern Generator Moku:Go Oscilloscope & Voltmeter Moku:Go PID Controller Moku:Go Spectrum Analyzer Moku:Go Waveform Generator Moku:Go Power Supplies Moku:Go Lock-in Amplifier Moku:Go Time & Frequency Analyzer Moku:Go Laser Lock Box Moku:Go Phasemeter
  • Moku:Lab
    Moku:Lab General Moku:Lab Arbitrary Waveform Generator Moku:Lab Data Logger Moku:Lab Digital Filter Box Moku:Lab FIR Filter Builder Moku:Lab Frequency Response Analyzer Moku:Lab Laser Lock Box Moku:Lab Lock-in Amplifier Moku:Lab Oscilloscope Moku:Lab Phasemeter Moku:Lab PID Controller Moku:Lab Spectrum Analyzer Moku:Lab Time & Frequency Analyzer Moku:Lab Waveform Generator Moku:Lab Logic Analyzer/Pattern Generator
  • Moku:Pro
    Moku:Pro General Moku:Pro Arbitrary Waveform Generator Moku:Pro Data Logger Moku:Pro Frequency Response Analyzer Moku:Pro Oscilloscope Moku:Pro PID Controller Moku:Pro Spectrum Analyzer Moku:Pro Waveform Generator Moku:Pro Lock-in Amplifier Moku:Pro Laser Lock Box Moku:Pro Digital Filter Box Moku:Pro FIR Filter Builder Moku:Pro Phasemeter Moku:Pro Multi-instrument Mode Moku:Pro Logic Analyzer/Pattern Generator Moku:Pro Time & Frequency Analyzer
  • Python API
  • MATLAB API
  • Arbitrary Waveform Generator
  • Data Logger
  • Digital Filter Box
  • FIR Filter Builder
  • Frequency Response Analyzer
  • Laser Lock Box
  • Lock-in Amplifier
  • Oscilloscope
  • Phasemeter
  • PID Controller
  • Spectrum Analyzer
  • Time & Frequency Analyzer
  • Waveform Generator
  • Logic Analyzer & Pattern Generator
  • Multi Instrument Mode
  • Moku Cloud Compile
  • Moku general
  • LabVIEW
+ More

When using Multi-Instrument Mode (MiM), there are four instruments that can cause the warning “Instrument frame rates may be reduced in this configuration” to appear when placed in one of the instrument slots. The four instruments are the Oscilloscope, Laser Lock Box, Lock-In Amplifier, and Logic Analyzer. These instruments can acquire data at the full sampling rate of the device. The reduced frame rate warning will appear if one or more of these four instruments are used in MiM simultaneously with one another. For example, two LIA's being used at once (as shown in the screenshot below), or a LIA with a LLB, etc. This warning will appear, however, the effect on the frame rate will only be observed if all of the following criteria are met:

  • the oscilloscope or an embedded oscilloscope is being used in one of the instruments mentioned above
  • the time span of the scope is small enough (zoomed in) so that the sampling rate is at its maximum
  • more than one of the scopes in MiM are operating at this maximum sampling rate
reduced rates performance warning

Was this article helpful?

Yes
No
Give feedback about this article

Related Articles

  • What's the difference between the spectrum analyzer and oscilloscope FFT?
  • Phasemeter output waveform resolution

Sitemap

  • Moku:Lab
  • Instruments
  • Software
  • Company
  • Support
  • Store
  • Terms & Conditions
  • Privacy Policy

Offices

United States
+1 (619) 332-6230
12526 High Bluff Dr
Suite 150
San Diego, CA 92130

Australia
+61 2 6171 9730
243 Northbourne Avenue
Suite 2
Lyneham, ACT 2602

Australia
+61 03 7073 3594
700 Swanston Street
Suite 5E, Level 5
Carlton, VIC 3053

Follow us

Youtube LinkedIn

官方微信

Contact us
© 2025 Liquid Instruments. All rights reserved.

Knowledge Base Software powered by Helpjuice

Definition by Author

0
0
Expand